T-15MD Tokamak Microwave Interferometer for Measuring the Average Electron Density of Plasma

A. S. Drozd,D. S. Sergeev, R. A. Begishev, G. B. Igon’kina,M. M. Sokolov, N. V. Korshunov, E. N. Khairutdinov,T. B. Myalton

Plasma Physics Reports(2024)

引用 0|浏览1
暂无评分
摘要
The distinctive features of the T-15MD tokamak microwave interferometer for measuring linearly integrated electron density, as well as the system for processing and recording its signals, are presented. The phase stability of microwave interferometer signals is analyzed. The results of measurements with a microwave interferometer during the first experimental campaign of the T-15MD tokamak are presented. The phase values were extracted by using an analog phase meter and post-processing of digitized microwave interferometer signals: an intermediate frequency signal and a local oscillator signal. It is shown that the results of the plasma density calculation by these two methods coincide.
更多
查看译文
关键词
microwave interferometer,tokamak,electron density,signal processing,plasma measurements