Two cases of silent subcutaneous implantable cardioverter defibrillator electrode displacement.

Pacing and clinical electrophysiology : PACE(2024)

Cited 0|Views2
No score
Abstract
We describe two cases of secondary prevention subcutaneous implantable cardioverter defibrillator (S-ICD) implantation and subsequent S-ICD electrode displacement which initially went undetected. One presentation was a result of a coincidental chest x-ray for respiratory exacerbation and another with an untreated episode highlighted via remote monitoring, both patients were booked to clinic for further investigation. Our findings highlighted had there been a comparison of the existing subcutaneous electrogram (S-ECG) to captured S-ECGs at time of implant the electrode displacement would have been detected beforehand. This underpins the importance of introducing the simple management strategy into routine follow-up.