Antimicrobial susceptibility testing (AST) by disk diffusion provides an accurate image of bacterial growth, enabling the detection of culture purity, heterogeneous growth, and antibiotic interactions. However, this manual method is time-consuming and visual interpretation is prone to errors. To overcome these disadvantages, the Radian® In-Line Carousel (Copan, Brescia, Italy) was launched, which is a WASPLab® module dedicated to full automation of (pre)-analytical steps as well as interpretation of disk diffusion AST. However, until now, no evaluation of Radian® against manual disk diffusion has been performed. We assessed the categorical agreement (CA) between standardized disk diffusion (reference method) and Radian® using EUCAST 2021 breakpoints. We tested 135 non-duplicate strains, selected from the National EUCAST challenge panel, clinical strains, and external quality controls. The strains included Enterobacterales (n = 63), Enterococcus faecalis (n = 3), Enterococcus faecium (n = 10), Pseudomonas aeruginosa (n = 16), Staphylococcus aureus (n = 19), coagulase-negative staphylococci (n = 4), and Streptococcus spp. (n = 20). Furthermore, we explored antibiotic disk thermolability in the WASP Radian® carousel by testing 10 ATCC® strains up to 7 days. The observed CA was 95.3