摘要
OBJECTIVES:Differentiation of iris and ciliary body lesions as benign or malignant and cystic or solid is important. The aim of this study was to compare anterior segment swept-source optical coherence tomography (AS SS-OCT) and ultrasound biomicroscopy (UBM) findings in iris and ciliary body tumors. RESEARCH DESIGN AND METHODS:Forty-two eyes of 38 cases with iris and ciliary body tumors imaged with UBM and AS SS-OCT between September 2018 and September 2023 were evaluated retrospectively. RESULTS:Of 42 eyes, 14 had melanoma, 14 iris pigment epithelial (IPE) cysts, 7 nevi, 3 Lisch nodules, 2 iris stromal cysts, 1 pars plana cysts, and 1 iris mammillations. An equivalent (100%) visualization of the anterior tumor margin was obtained with both techniques. Compared to AS SS-OCT, UBM was superior for posterior margin visualization in melanocytic tumors and IPE cysts. Bland-Altman plots demonstrated good agreement between UBM and AS SS-OCT for melanocytic tumors < 2.5 mm in base diameter and < 2 mm in thickness. CONCLUSIONS:Although, UBM is the gold standard for ciliary body and iridociliary tumors. AS SS-OCT should be considered as an excellent alternative to UBM, especially in minimally elevated iris lesions.