Impact of beam complexity on plan delivery accuracy verification of a transmission detector in volumetric modulated arc therapy

Physica Medica(2024)

引用 0|浏览3
暂无评分
摘要
Objective To study the effect of beam complexity on VMAT delivery accuracy evaluated by means of a transmission detector, together with the possibility of scoring plan complexity. Methods 43 clinical VMAT plans delivered by a TrueBeam linear accelerator to both Delta4 Discover and Delta4 Phantom+ for patient-specific quality assurance were evaluated. Global Dose-γ analysis, MLC-γ analysis, percentage of leaves with a deviation between planned and measured leaf tip position lower than 1 mm (LD) were computed. Modulation complexity score (MCSv), average leaf travel (LT), a multiplicative combination of LT and MCSv (LTMCS), percentage of leaves with speed lower than 5 mm/s (LS), from 5 to 20 mm/s (MS), higher than 20 mm/s (HS) and the average value of leaf speed (MLCSav) were evaluated by means of an home-made Matlab script. Results Dose-γ passing rate showed a moderate correlation with MCSv, LT, MLCSav, LS and HS, while a stronger positive correlation was found with LTMCS. A strong correlation was observed between LD and both LT and leaves speed, while a weak correlation was observed with MCSv. A correlation between MLC-γ pass rate and plan complexity parameters was found except for MCSv; a moderate correlation with LS was observed, while all other parameters showed weak correlations. Conclusions The study confirmed the possibility to establish correlations between plan complexity indices versus dose distribution and MLC parameters measured by a transmissive detector. Further investigation is necessary to define specific values of the complexity indices to evaluate whether a VMAT plan is deliverable as intended.
更多
查看译文
关键词
Transmission detector,Volumetric modulated arc therapy,Complexity plan indices
AI 理解论文
溯源树
样例
生成溯源树,研究论文发展脉络
Chat Paper
正在生成论文摘要