Measurement of the response of a 6Li-glass detector to gamma rays by a coincidence method

Fumiaki Ito,Jaehong Lee, Kota Hironaka,Tone Takahashi, Satoshi Suzuki, Takanori Mochimaru,Jun-ichi Hori,Kazushi Terada,Mitsuo Koizumi

Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment(2024)

Cited 0|Views2
No score
Abstract
The response of a gamma-ray spectrometer is generally determined by analyzing full-energy peaks. However, full-energy peaks cannot be measured easily in the case of scintillation detectors that consist of light elements, such as glass scintillators. Only a strong Compton plateau appears in the spectrum of such detectors. Therefore, Compton edgers were used to evaluate the response of these detectors. The response of a low-resolution 6Li-glass detector to gamma rays was measured for the first time by a coincidence method with a high-resolution LaBr3:Ce detector using cascade gamma rays (2.75 and 1.37 MeV) from a 24Na source. Coincidence gates were applied at the peaks of the spectrum of the LaBr3:Ce detector at the 0.51 MeV annihilation peak, and the sum peaks of a gamma ray and a backscattered gamma ray. By analyzing the gated spectra of the 6Li-glass detector, the energy-dependent detector response (i.e., the output strength and its dispersion) was determined.
More
Translated text
Key words
6Li-glass,Scintillator,GS20,Neutron,Gamma ray,Response characteristics,Energy resolution,Coincidence method,LaBr3:Ce,Cascade gamma-ray emission,24Na