Innovations in dedicated PET instrumentation: from the operating room to specimen imaging.

Physics in medicine and biology(2024)

Cited 0|Views9
No score
Abstract
This review casts a spotlight on intraoperative PET scanners and the distinctive challenges they confront. Specifically, these systems contend with the necessity of partial coverage geometry, essential for ensuring adequate access to the patient. This inherently leans them towards limited-angle PET imaging, bringing along its array of reconstruction and geometrical sensitivity challenges. Compounding this, the need for real-time imaging in navigation systems mandates rapid acquisition and reconstruction times. For these systems, the emphasis is on dependable PET image reconstruction (without significant artifacts) while rapid processing takes precedence over the spatial resolution of the system. In contrast, specimen PET imagers are unburdened by the geometrical sensitivity challenges, thanks to their ability to leverage full coverage PET imaging geometries. For these devices, the focus shifts: high spatial resolution imaging takes precedence over rapid image reconstruction. This review concurrently probes into the technical complexities of both intraoperative and specimen PET imaging, shedding light on their recent designs, inherent challenges, and technological advancements.