Illumination matters Part II: advanced comparative analysis of flexible ureteroscopes in a kidney model by PEARLS

WORLD JOURNAL OF UROLOGY(2024)

引用 0|浏览3
暂无评分
摘要
Purpose The aim of the study was to evaluate illumination properties in an in-vitro kidney calyx model in saline. Design and methods We evaluated a series of contemporary flexible ureteroscopes including the Storz Flex-Xc and Flex-X2s, Olympus V3 and P7, Pusen 7.5F and 9.2F, as well as OTU WiScope using a 3D-printed closed pink kidney calyx model, submerged in saline. A spectrometer was used for illuminance and color temperature measurements at different openings located at center (direct light), 45 degrees (direct and indirect light) and 90 degrees(indirect light) to the axis of the scope. Results Maximum illuminance was at the center opening for all scopes (range: 284 to 12,058 lx at 50% brightness and 454 to 11,871 lx at 100% brightness settings). The scope with the highest center illuminance (Flex-Xc) was 26 times superior to the scope with the lowest illuminance (Pusen 7.5Fr) at 100% brightness setting. For each scope, there was a peripheral illuminance drop ranging from - 43 to - 92% at 50% brightness and - 43% to - 88% at 100% brightness settings, respectively (all p < 0.01). Highest drop was for the P7 and the Pusen 9.2F. All scopes had illuminance skew, except the V3. All scopes had a warm color temperature. Conclusion Illumination properties vary between ureteroscopes in an enclosed cavity in saline, and differs at center vs 45 degrees and 90 degrees positions within scopes. Peripheral illuminance drop can be as high as - 92%, which is undesirable. This may affect the choice of ureteroscope and light brightness settings used in surgery by urologists.
更多
查看译文
关键词
Flexible ureteroscopy,Light distribution,Illumination properties,Illuminance,Saline,Enclosed cavity
AI 理解论文
溯源树
样例
生成溯源树,研究论文发展脉络
Chat Paper
正在生成论文摘要