Diagnostic Accuracy of the Abbott SD Bioline MPT64 Antigen Test for Identification of MTB Complex in a U.S. Clinical Mycobacteriology Laboratory

D.T. Armstrong, Logan Pretty, Karen D’Agostino, Roxanne Redhead-Harper,Nicole Parrish

Heliyon(2024)

引用 0|浏览1
暂无评分
摘要
Identification of the Mycobacterium tuberculosis complex (MTBC) from culture and differentiation from other non-tuberculous mycobacterial species is required for rapid diagnosis and accurate treatment. However, gaps exist in culture-based identification of acid-fast bacilli (AFB) positive cultures for rapid rule-out of MTBC in the United States. The SD Bioline™ MPT64 (Abbott Inc, South Korea) lateral flow assay (LFA) has high sensitivity and specificity for the detection of MTBC in liquid culture but has not been evaluated in a clinical mycobacteriology laboratory in the United States. We conducted a diagnostic accuracy study of this LFA for detection of MTBC versus NTMs on AFB positive cultures. A total of 362 tests were performed, with a sensitivity and specificity of 100% (362/362) across all tests. The SD Bioline MPT64 assay provides accurate test results with AFB-positive liquid cultures and could fill the current gap for rapid rule-out of MTBC in U.S.-based clinical laboratories.
更多
查看译文
关键词
MPT64,antigen,culture,mycobacteriology,tuberculosis
AI 理解论文
溯源树
样例
生成溯源树,研究论文发展脉络
Chat Paper
正在生成论文摘要