The Impact of Repeated Steam Sterilization Cycles on the Efficacy of Chairside Adjustment Kits for Polishing Monolithic Multi-Layered Zirconia Dental Restoration Material

Mohammed E. Sayed,Honey Lunkad,Enas Jaber Mohana, Sara Ahmad Mesawa, Fatimah Ibrahim Dowairi, Emad Aldeen A. Alharbi, Hayat Ali Hakami,Hossam F. Jokhadar,Nasser M. Alqahtani,Abdullah Hasan Alshehri,Ebrahim Fihaid Alsubaiy,Saeed M. Alqahtani, Nagam Mohammed Mohsen, Mohammed H. D. Al Wadei,Khurshid Mattoo

Medical science monitor(2024)

Cited 0|Views9
No score
Abstract
BACKGROUND Before insertion, chairside adjustment kits are heat sterilized for positioning and polishing dental restorations. This study aimed to evaluate the effects of 2 steam sterilization cycles on the efficacy of polishing highly translucent monolithic zirconia (HTMLZ) dental restoration material. MATERIAL AND METHODS 100 HTMLZ disc-shaped specimens were adjusted (grinding, finishing, polishing) with EVE Diacera kit. Two steam sterilization techniques [standard (Gp S), immediate/flash (Gp (F)] of CAK were further subgrouped based on number of sterilization cycles [cycle 1 (control), cycle 5, 10, 15, and 20 (experimental)] (n=10 each). Each subgroup accordingly was evaluated for average surface roughness (Ra) and root mean square roughness (Rq) using a profilometer. Mean and standard deviation of 5 subgroups were statistically analyzed using one-way ANOVA/post hoc Tukey's test. Scanning electron microscopy complemented Ra, Rq measurements. Statistical differences of P≤0.05 were considered significant. RESULTS HTMLZ specimens in both groups showed increased (Ra/Rq) values after repeated sterilization of EVE Diacera kit, with Gp F showing lesser increase than Gp S (20 cycles). Gp F at 10 cycles and Gp S at 15 cycles showed clinically unacceptable roughness threshold (0.25 μm). Differences between subgroups for Ra and Rq values were significant (P≤0.05) with less differences within groups observed in early cycles (1, 10). Results validate the manufacturer's recommendations of using flash sterilization/10 cycles for EVE Diacera kit. CONCLUSIONS Repeated sterilization reduces efficacy of chairside adjustment kit to produce smooth surfaces on HTMLZ. This study recommends flash sterilization to a maximum of 10 times to get the clinically acceptable results of Ra and Rq.
More
Translated text
Key words
Disinfection,Steam,Sterilization,Yttria Stabilized Tetragonal Zirconia,Denture, Partial, Fixed
AI Read Science
Must-Reading Tree
Example
Generate MRT to find the research sequence of this paper
Chat Paper
Summary is being generated by the instructions you defined