Characterisation of Redlen HF-CdZnTe at > 106 ph s-1 mm-2 using HEXITECMHz

B.D. Cline, D. Banks, S. Bell, I. Church, A. Davis, T. Gardiner, J. Harris, M. Hart,L. Jones, T. Nicholls, J. Nobes, S. Pradeep, M. Roberts, D. Sole, M.C. Veale, M.D. Wilson, V. Dhamgaye, O. Fox, K. Sawhney

Journal of Instrumentation(2024)

引用 0|浏览0
暂无评分
摘要
Abstract In this paper, results are presented from the characterisation of Redlen Technologies high-flux-capable Cadmium Zinc Telluride (HF-CZT) hybridised to the HEXITECMHz ASIC, a novel 1 MHz continuous X-ray imaging system. A 2 mm thick HF-CZT HEXITECMHz detector was characterised on the B16 Test Beamline at the Diamond Light Source and displayed an average FWHM of 850 eV for monochromatic X-rays of energy 20 keV. Measurements revealed a shift in the baseline of irradiated pixels that results in a movement of the entire spectrum to higher ADU values. Datasets taken to analyse the effect's dynamics showed it to be highly localised and flux-dependent, with the excess leakage current generated equivalent to per-pixel shifts of ∼ 543 pA (8.68 nA mm-2) at a flux of 1.26×107 ph s-1 mm-2. Comparison to results from a p-type Si HEXITECMHz device indicate this `excess leakage-current' effect is unique to HF-CZT and it is hypothesised that it originates from trapping at the electrode-CZT interface and a temporary modification of the potential barrier between the CZT and metal electrode.