The continuous exposure of blood to a non-biological surface during extracorporeal membrane oxygenation (ECMO) may lead to progressive thrombus formation in the oxygenator, hemolysis and consequently impaired gas exchange. In most centers oxygenator performance is monitored only on a once daily basis. Carboxyhemoglobin (COHb) is generated upon red cell lysis and is routinely measured with any co-oximetry performed to surveille gas exchange and acid–base homeostasis every couple of hours. This retrospective cohort study aims to evaluate COHb in the arterial blood gas as a novel marker of oxygenator dysfunction and its predictive value for imminent oxygenator change. Out of the 484 screened patients on ECMO 89, cumulatively requiring 116 oxygenator changes within 1833 patient days, including 19,692 arterial COHb measurements were analyzed. Higher COHb levels were associated with lower post-oxygenator pO2 (estimate for log(COHb): − 2.176 [95