摘要
Pulse oximeters’ (POs) varying performance based on skin tones has been highly publicised. Compared to arterial blood gas analysis, POs tend to overestimate oxygen saturation (SpO2) values for people with darker skin (occult hypoxemia). The objective is to develop a test bench for assessing commercial home and hospital-based POs in controlled laboratory conditions. A laboratory simulator was used to mimic different SpO2 values ( 70 to 100
更多
查看译文
关键词
Pulse oximeter,Oxygen saturation,Transmission-mode,Occult hypoxemia,Racial bias,Melanin,Skin colour