Repeatability assessment of anterior segment measurements in myopic patients using an anterior segment OCT with placido corneal topography and agreement with a swept-source OCT

Hao Wang, Li-Shuang Zhu,Chen-Jiu Pang,Qi Fan

BMC Ophthalmology(2024)

引用 0|浏览0
暂无评分
摘要
The precision of anterior segment biometric measurements in eyes has become increasingly important in refractive surgery. The purpose of this study is to assess the repeatability of the automatic measurements provided by a new spectral-domain optical coherence tomograph (SD-OCT)/Placido topographer (MS-39, CSO) and its agreement with a swept-source OCT (SS-OCT) biometer (CASIA SS-1000, Tomey) in patients with myopia. The right eye of 235 subjects was scanned 3 times with both devices. The evaluated parameters included central corneal radius of the steep meridian, central corneal radius of the flat meridian, mean central corneal radius, thinnest corneal thickness, central corneal thickness, anterior chamber depth, corneal volume and diameter. The intraobserver repeatability of the MS-39 measurements was calculated using intraclass correlation coefficient (ICC), within subject standard deviation, coefficient of repeatability, coefficient of variation and repeated-measures analysis of variance of the 3 repeated measurements. The agreement between the two devices was evaluated by 95
更多
查看译文
关键词
OCT,Corneal thickness,Keratometry,Corneal volume,Repeatability,Agreement