Mechanical failure of distal femur mega prosthesis due to polyaryl-ether-ether-ketone (PEEK) hinge component

Archives of Orthopaedic and Trauma Surgery(2024)

引用 0|浏览0
暂无评分
摘要
Polyaryl-ether-ether-ketone (PEEK) has gained popularity as a substrate for orthopaedic hardware due to its desirable properties such as heat and deformation resistance, low weight, and ease of manufacturing. However, we observed a relatively high failure rate of PEEK-based hinges in a distal femur reconstruction system. In this study, we aimed to evaluate the proportion of patients who experienced implant failure, analyse the mechanism of failure, and document the associated clinical findings. We conducted a retrospective cohort study, reviewing the medical charts of 56 patients who underwent distal femur resection and reconstruction with a PEEK Optima hinge-based prosthesis between 2004 and 2018. Concurrently, we performed a clinical and biomechanical failure analysis. PEEK component failure occurred in 21 out of 56 patients (37.5
更多
查看译文
关键词
Polyaryl-ether-ether-ketone,PEEK,Biomechanical failure,Prostheses,Femur reconstruction