A Pneumatic Driven MRI-Guided Robot System for Prostate Interventions

IEEE Transactions on Medical Robotics and Bionics(2024)

Cited 0|Views6
No score
Abstract
Under the guidance of high-resolution Magnetic Resonance Imaging (MRI), robotic devices offer a great advantage for prostate intervention. This paper presents an MR-safe robot, where a needle is attached to the needle guide to obtain prostate biopsies during surgeries. The robot is powered by three actuators, two of them are customized to function as a work plane that allows the needle to move horizontally and vertically, and the third actuator controls the rotation of the work plane, allowing the needle to be inserted into the prostate from different directions. All the actuators are pneumatically actuated to allow them to work in a Magnetic Resonance (MR) environment. The kinematics and mechanism of the robot are analyzed. A user interface developed using LabView is created to calculate the target position and generate a control signal for the valves. In the open-air test, the needle can reach the target with an accuracy of 1.3 mm. The signal-to-noise ratio (SNR) variation was measured below 5% under a 3T MR scanner.
More
Translated text
Key words
MRI,Prostate biopsy,MR-safe robot