Optimizing cone-beam computed tomography exposure for an effective radiation dose and image quality balance

IMAGING SCIENCE IN DENTISTRY(2024)

引用 0|浏览3
暂无评分
摘要
Purpose: The aim of this study was to evaluate the influence of different cone-beam computed tomography (CBCT) acquisition protocols on reducing the effective radiation dose while maintaining image quality. Materials and Methods: The effective dose emitted by a CBCT device was calculated using thermoluminescent dosimeters placed in a Rando Alderson phantom. Image quality was assessed by 3 experienced evaluators. The relationship between image quality and confidence was evaluated using the Fisher exact test, and the agreement among raters was assessed using the kappa test. Multiple linear regression analysis was performed to investigate whether the technical parameters could predict the effective dose. P-values<0.05 were considered to indicate statistical significance. Results: The optimized protocol (3 mA, 99 kVp, and 450 projection images) demonstrated good image quality and a lower effective dose for radiation-sensitive organs. Image quality and confidence had consistent values for all structures (P<0.05). Multiple linear regression analysis resulted in a statistically significant model. The milliamperage (b = 0.504; t = 3.406; P = 0.027), kilovoltage peak (b = 0.589; t = 3.979; P = 0.016) and number of projection images (b = 0.557; t = 3.762; P = 0.020) were predictors of the effective dose. Conclusion: Optimized CBCT acquisition protocols can significantly reduce the effective radiation dose while maintaining acceptable image quality by adjusting the milliamperage and projection images. (Imaging Sci Dent 20230251)
更多
查看译文
关键词
Cone-Beam Computed Tomography,Radiation Exposure,Radiation Dosimeters
AI 理解论文
溯源树
样例
生成溯源树,研究论文发展脉络
Chat Paper
正在生成论文摘要