Long-Term Ultrasound Twinkling Detectability and Safety of a Polymethyl Methacrylate Soft Tissue Marker Compared to Conventional Breast Biopsy Markers—A Preclinical Study in a Porcine Model

Christine U. Lee,Matthew W. Urban,Gina K. Hesley,Benjamin G. Wood, Thomas R. Meier,Beiyun Chen, Blake A. Kassmeyer, Nicholas B. Larson,A. Lee Miller, James L. Herrick,James W. Jakub,Mara A. Piltin

Ultrasound in Medicine & Biology(2024)

引用 0|浏览0
暂无评分
摘要
Objective We have studied the use of polymethyl methacrylate (PMMA) as an alternative biopsy marker that is readily detectable with ultrasound Doppler twinkling in cases of in vitro, ex vivo, or limited duration in vivo settings. This study investigates the long-term safety and ultrasound Doppler twinkling detectability of a PMMA breast biopsy marker following local perturbations and different dwell times in a 6-mo animal experiment. Methods This study, which was approved by our Institutional Animal Care and Use Committee, involved three pigs and utilized various markers, including PMMA (Zimmer Biomet), 3D-printed, and Tumark Q markers. Markers were implanted at different times for each pig. Mesh material or ethanol was used to induce a local inflammatory reaction near certain markers. A semiquantitative twinkling score assessed twinkling for actionable localization during monthly ultrasounds. At the primary endpoint, ultrasound-guided localization of lymph nodes with detectable markers was performed. Following surgical resection of the localized nodes, histomorphometric analysis was conducted to evaluate for tissue ingrowth and the formation of a tissue rind around the markers. Results No adverse events occurred. Twinkling scores of all markers for all three pigs decreased gradually over time. The Q marker exhibited the highest mean twinkling score followed by the PMMA marker, PMMA with mesh, and Q with ethanol. The 3D-printed marker with mesh and PMMA with ethanol had the lowest scores. All wire-localized lymph nodes were successfully resected. Despite varying percentages of tissue rind around the markers and a significant reduction in overall twinkling (p < 0.001) over time, mean PMMA twinkling scores remained clinically actionable at 6 and 5 mo using a General Electric C1-6 probe and 9L-probe, respectively. Conclusions In this porcine model, the PMMA marker demonstrates an acceptable safety profile. Clinically actionable twinkling aids PMMA marker detection even after 6 mo of dwell time in porcine lymph nodes. The Q marker maintained the greatest twinkling over time compared to all the other markers studied.
更多
查看译文
关键词
Ultrasound color Doppler,Ultrasound,Twinkling,Polymethyl methacrylate,Bone cement,PMMA,Breast biopsy marker,Localization,Breast cancer,Neoadjuvant therapy
AI 理解论文
溯源树
样例
生成溯源树,研究论文发展脉络
Chat Paper
正在生成论文摘要