Extended just-noticeable difference for ultralow-luminance displays used in diagnostic imaging
JOURNAL OF THE OPTICAL SOCIETY OF AMERICA A-OPTICS IMAGE SCIENCE AND VISION(2024)
Abstract
This study investigates the just-noticeable difference (JND) in luminance below the 0.050 cd/m2 threshold not covered by the DICOM standard in medical imaging displays. A total of 21 healthy young adults were tested using an ultralow-luminance liquid crystal display at a viewing distance of 46 cm. The optotype featured a 4 cycle/deg rectangular wave-like stripe. The average percentage of correct responses for JND indices between -18 and 0 was 70.8%. Using the extended JND based on the Barten model that matches the current JND definition, we confirmed the JND at a very low luminance of 0.0500 cd/m2 or less. These findings suggest the feasibility of further refining the gradation differences in medical imaging displays. (c) 2024 Optica Publishing Group