Evaluation of effects from hemoglobin variants on HbA1c measurements by different methods

CLINICAL CHEMISTRY AND LABORATORY MEDICINE(2024)

引用 0|浏览2
暂无评分
摘要
Objectives: The impact of seven hemoglobin variants (Hb Q-Thailand, Hb G-Honolulu, Hb Ube-2, Hb New York, Hb J-Bangkok, Hb G-Coushatta, and Hb E) on the outcome of HbA(1c )was investigated for six methods by comparing with liquid chromatography-tandem mass spectrometry (LC/MS/MS) reference method. Methods: Twenty-nine normal and 112 variant samples were measured by LC/MS/MS, Sebia Capillarys 3 TERA, Intelligene Biosystems QuanTOF, Premier Hb9210, Arkray HA-8190V, Bio-Rad D-100, and Tosoh G11, then evaluated for correlation, consistency, and mean relative bias among six methods. The lowest biological variation bias of +/- 2.8 % was an acceptable standard. Results: All methods showed poor correlation and consistency with LC/MS/MS for Hb E. The unacceptable biases were observed for Capillarys 3 TERA (-14.4 to -3.7 % for Hb Q-Thailand, Hb Ube-2, Hb New York, Hb J-Bangkok and Hb E), QuanTOF (-8.3 to -2.9 % for Hb Ube-2, Hb New York and Hb G-Coushatta), Premier Hb9210 (-18.3 to -3.6 % for Hb Q-Thailand, Hb Ube-2, Hb New York, Hb J-Bangkok and Hb E), HA-8190V variant mode (-17.3 to 6.6 % for Hb G-Honolulu, Hb Ube-2, Hb New York, Hb G-Coushatta and Hb E). All variant samples showed larger biases than +/- 2.8 % comparing HA-8190V fast mode, D-100, and G11 with LC/MS/MS. Conclusions: The accuracy of different HbA(1c) methods was influenced by some Hb variants, especially Hb Ube-2 and Hb New York. Thus, laboratories need to choose appropriate methods to measure HbA1c with different Hb variants.
更多
查看译文
关键词
HbA(1c),hemoglobin variants,method comparison
AI 理解论文
溯源树
样例
生成溯源树,研究论文发展脉络
Chat Paper
正在生成论文摘要