Low-dose CT fluoroscopy-guided interventional minimally invasive robot

HELIYON(2024)

引用 0|浏览7
暂无评分
摘要
Background: This study aimed to assess the feasibility, safety, and accuracy of a low-dose CT fluoroscopy-guided remote-controlled robotic real -time puncture procedure. Methods: The study involved two control groups with Taguchi method: Group A, which underwent low-dose traditional CT-guided manual puncture (blank control), and Group B, which underwent conditional control puncture. Additionally, an experimental group, Group C, underwent CT fluoroscopy-guided remote-controlled robotic real -time puncture. In a phantom experiment, various simulated targets were punctured, while in an animal experiment, attempts were made to puncture targets in different organs of four pigs. The number of needle adjustments, puncture time, total puncture operation time, and radiation dose were analyzed to evaluate the robot system. Results: Successful punctures were achieved for each target, and no complications were observed. Dates were calculated for all parameters using Taguchi method. Conclusion: The low-dose CT fluoroscopy-guided puncture robot system is a safe, feasible, and equally accurate alternative to traditional manual puncture procedures.
更多
查看译文
关键词
CT fluoroscopy,Low-dose,Robot-assisted,Teleoperation
AI 理解论文
溯源树
样例
生成溯源树,研究论文发展脉络
Chat Paper
正在生成论文摘要