An automated fluorescent immunoassay for on-site screening of AFM1 in raw milk at the ppt level

One Health Advances(2024)

引用 0|浏览10
暂无评分
摘要
In the dairy industry, the quality of raw milk as it is collected from dairy farmers must be strictly ensured. Therefore, when on-site screening of typical carcinogens in raw milk, this technology must simultaneously be convenient to operate and highly sensitive. Here, an automated and fluorescent immunoassay system for screening trace amounts of aflatoxin M1 (AFM1) in raw milk was developed. The whole immunoassay procedure can be processed in a true “sample-to-results” paradigm, avoiding the tedious manual operation of the traditional indirect competitive enzyme-linked immunosorbent assay (ic-ELISA) method. In addition, we designed an integrated fluorescent spectrometer that can quantitatively measure fluorescent signals with high sensitivity. The automated and fluorescent immunoassay system can screen AFM1 in raw milk samples with an ultra low limit of detection (4.7 pg/mL in raw milk). The half maximal inhibitory concentration (IC50) was 8.3 pg/mL, which is approximately 4-fold lower than that of traditional ic-ELISA. When the system was applied to screen actual raw milk samples, the recovery rates ranged from 79.4
更多
查看译文
关键词
AFM1,An automated immunoassay,An integrated fluorescent detector,Raw milk,Contaminants screening