Computational analysis and experimental validation of dialysis flow rate in central venous catheters

Journal of the Brazilian Society of Mechanical Sciences and Engineering(2024)

引用 0|浏览2
暂无评分
摘要
Despite a large number of studies on the favorable correlation between the higher dialysis dose and clinical efficiency of hemodialysis treatment, the effect of the dialysis flow rate inside the catheter has been little investigated from the perspective of hemodynamic modifications in the central venous access for hemodialysis. This study proposes the computational analysis and an in vitro experimental validation of this model to assess hemodynamic behavior and the thrombogenic potential in central venous access for hemodialysis, using different flow rates in the central venous catheter (200, 250, 300, 350 and 400 mL/min). A mathematical model based on the Eulerian formulation was implemented in the solver subroutine to calculate the linear platelet lysis index (PLIL). The pressure drop obtained from the computational and experimental models was compared to validate the computational model. The maximum percentage difference between these models was less than 13
更多
查看译文
关键词
Blood flow rate,Central venous catheter,Thrombus formation,Hemodialysis,Computational fluid dynamics,Computational model validation