Validation of image-derived input function using a long axial field of view PET/CT scanner for two different tracers

EJNMMI Physics(2024)

引用 0|浏览3
暂无评分
摘要
Accurate image-derived input function (IDIF) from highly sensitive large axial field of view (LAFOV) PET/CT scanners could avoid the need of invasive blood sampling for kinetic modelling. The aim is to validate the use of IDIF for two kinds of tracers, 3 different IDIF locations and 9 different reconstruction settings. Eight [18F]FDG and 10 [18F]DPA-714 scans were acquired respectively during 70 and 60 min on the Vision Quadra PET/CT system. PET images were reconstructed using various reconstruction settings. IDIFs were taken from ascending aorta (AA), descending aorta (DA), and left ventricular cavity (LV). The calibration factor (CF) extracted from the comparison between the IDIFs and the manual blood samples as reference was used for IDIFs accuracy and precision assessment. To illustrate the effect of various calibrated-IDIFs on Patlak linearization for [18F]FDG and Logan linearization for [18F]DPA-714, the same target time-activity curves were applied for each calibrated-IDIF. For [18F]FDG, the accuracy and precision of the IDIFs were high (mean CF ≥ 0.82, SD ≤ 0.06). Compared to the striatum influx (Ki) extracted using calibrated AA IDIF with the updated European Association of Nuclear Medicine Research Ltd. standard reconstruction (EARL2), Ki mean differences were < 2
更多
查看译文
关键词
[18F]DPA-714,[18F]FDG,LAFOV PET/CT,IDIF