Interaction Between Microbubbles and Microwave Ablation: A Phantom and Rabbit Model.

Journal of ultrasound in medicine : official journal of the American Institute of Ultrasound in Medicine(2024)

引用 0|浏览2
暂无评分
摘要
OBJECTIVES:This study aimed to explore the interactions between microbubbles and microwave ablation (MWA). METHODS:The study employed custom-made phantoms (in vitro) and white New Zealand rabbits (in vivo). MWA was performed with or without microbubbles in the phantoms (2 × 105 particles mL-1) and rabbit livers (intravenous injection of 0.05 mL/kg SonoVue). During the MWA, K-type thermocouple probes were used to monitor the MWA-induced temperature increase. Contrast-enhanced ultrasound imaging (CEUS) was used to monitor and analyze the microbubbles signal intensity. After MWA, the ablation-zone volumes were evaluated and compared between the groups with and without microbubbles. RESULTS:In both the phantom models and rabbits, microbubbles showed no significant influence on MWA, including the ablation range and MWA-induced temperature increase. In phantoms and rabbit livers filled with microbubbles, MWA caused the formation of a gradually expanding microbubble-defect region over the ablation time. An increase in the temperature caused microbubble destruction. CONCLUSIONS:Microbubbles had no significant influence on MWA. However, MWA induced the destruction of microbubbles in a temperature-dependent manner. Thus, the poor thermotolerance of microbubbles is a non-negligible barrier when using CEUS to monitor the ablation range during MWA in real-time.