Developments on Frequency Domain Multiplexing Readout for Large Arrays of Transition-Edge Sensor X-ray Micro-calorimeters

Journal of Low Temperature Physics(2024)

引用 0|浏览5
暂无评分
摘要
At SRON, we have been developing X-ray TES micro-calorimeters as backup technology for the X-ray Integral Field Unit (X-IFU) of the Athena mission, demonstrating excellent resolving powers both under DC and AC bias. We also developed a frequency-domain multiplexing (FDM) readout technology, where each TES is coupled to a superconducting band-pass LC resonator and AC biased at MHz frequencies through a common readout line. The TES signals are summed at the input of a superconducting quantum interference device (SQUID), which performs a first amplification at cryogenic stage. Custom analog front-end electronics and digital boards take care of further amplifying the signals at room temperature and of the modulation/demodulation of the TES signals and bias carrier, respectively. We report on the most recent developments on our FDM technology, which involves a two-channel demonstration with a total of 70 pixels with a summed energy resolution of 2.34 ± 0.02 eV at 5.9 keV without spectral performance degradation with respect to single-channel operation. Moreover, we discuss prospects towards the scaling-up to a larger multiplexing factor up to 78 pixels per channel in a 1–6 MHz readout bandwidth.
更多
查看译文
关键词
X-ray astronomy,Transition-edge sensors,Multiplexing
AI 理解论文
溯源树
样例
生成溯源树,研究论文发展脉络
Chat Paper