Osteoneogenesis at the Round Window: A Possible Cause of Cochlear Implant Failure?

AUDIOLOGY RESEARCH(2024)

引用 0|浏览2
暂无评分
摘要
Surgery for cochlear implant is a traumatic procedure, with inflammatory responses leading to immediate and delayed intracochlear changes, resulting in newly formed fibrous and bony tissue. This newly formed tissue is thought to affect speech perception with cochlear implants and can also play a role in causing device malfunctioning and soft failures. We present a case of left cochlear implant explantation and reimplantation in a 15-year-old girl, who experienced deterioration of speech perception and device failure associated with osteoneogenesis of the round window, which could represent a cause of cochlear implant failure. To avoid surgical trauma of the cochlear lateral wall, enlarged round window insertion rather than a cochleostomy, soft surgical techniques, and the application of steroids are all important issues to prevent new tissue formation, although special attention should also be given to the trauma of round window borders.
更多
查看译文
关键词
osteoneogenesis,round window,cochlear implant,device failure