Distances and angles in standing long-leg radiographs: comparing conventional radiography, digital radiography, and EOS

Skeletal Radiology(2024)

引用 0|浏览0
暂无评分
摘要
Distances and angles measured from long-leg radiographs (LLR) are important for surgical decision-making. However, projectional radiography suffers from distortion, potentially generating differences between measurement and true anatomical dimension. These phenomena are not uniform between conventional radiography (CR) digital radiography (DR) and fan-beam technology (EOS). We aimed to identify differences between these modalities in an experimental setup. A hemiskeleton was stabilized using an external fixator in neutral, valgus and varus knee alignment. Ten images were acquired for each alignment and each modality: one CR setup, two different DR systems, and an EOS. A total of 1680 measurements were acquired and analyzed. We observed great differences for dimensions and angles between the 4 modalities. Femoral head diameter measurements varied in the range of > 5 mm depending on the modality, with EOS being the closest to the true anatomical dimension. With functional leg length, a difference of 8.7
更多
查看译文
关键词
Conventional radiography,Digital radiography,Fan-beam,Cone-beam,Slot-scanner,EOS,Lower extremity,Varus,Valgus,Projectional distortion,Distortion