Real-Time Seizure Detection Using Behind-the-Ear Wearable System.
Journal of clinical neurophysiology : official publication of the American Electroencephalographic Society(2024)
Abstract
INTRODUCTION:This study examines the usability and comfort of a behind-the-ear seizure detection device called brain seizure detection (BrainSD) that captures ictal electroencephalogram (EEG) data using four scalp electrodes.
METHODS:This is a feasibility study. Thirty-two patients admitted to a level 4 Epilepsy Monitoring Unit were enrolled. The subjects wore BrainSD and the standard 21-channel video-EEG simultaneously. Epileptologists analyzed the EEG signals collected by BrainSD and validated it using video-EEG data to confirm its accuracy. A poststudy survey was completed by each participant to evaluate the comfort and usability of the device. In addition, a focus group of UT Southwestern epileptologists was held to discuss the features they would like to see in a home EEG-based seizure detection device such as BrainSD.
RESULTS:In total, BrainSD captured 11 of the 14 seizures that occurred while the device was being worn. All 11 seizures captured on BrainSD had focal onset, with three becoming bilateral tonic-clonic and one seizure being of subclinical status. The device was worn for an average of 41 hours. The poststudy survey showed that most users found the device comfortable, easy-to-use, and stated they would be interested in using BrainSD. Epileptologists in the focus group expressed a similar interest in BrainSD.
CONCLUSIONS:Brain seizure detection is able to detect EEG signals using four behind-the-ear electrodes. Its comfort, ease-of-use, and ability to detect numerous types of seizures make BrainSD an acceptable at-home EEG detection device from both the patient and provider perspective.
MoreTranslated text
AI Read Science
Must-Reading Tree
Example
![](https://originalfileserver.aminer.cn/sys/aminer/pubs/mrt_preview.jpeg)
Generate MRT to find the research sequence of this paper
Chat Paper
Summary is being generated by the instructions you defined