Quantification of chlorinated paraffins by chromatography coupled to high-resolution mass spectrometry - Part A: Influence of gas chromatography and ionisation source parameters.

Chemosphere(2024)

引用 0|浏览5
暂无评分
摘要
The analysis of chlorinated paraffins (CPs) has become a major analytical challenge. GC-ECNI-HRMS coupling is often used to analyse and quantify them. However, the influence of certain GC and ECNI parameters on the responses of polychlorinated n-alkanes (PCAs), the dominant components of CPs, has hardly been studied. In this paper, we investigated not only the influence of GC column characteristics, but also oven, GC inlet and source temperatures for simultaneous analysis of PCAs with chain-length ranging from 10 up to 20 carbon atoms (PCAs-C10-20). Particular attention was paid to the absolute response and PCA homologue group pattern obtained for a CP technical mixture. The optimum conditions for a wide homologue group determination were GC inlet, final gradient and ion source temperatures set at 220-240 °C, 340 °C and 200 °C. At the same time, a higher response was obtained with the Optima 5HT column compared to Optima 1 column, and with a length and film thickness of 12.5 m and 0.25 μm, respectively. The homologue group pattern of the technical mixture studied was significantly modified as a function of the source and GC inlet temperatures, film thickness and composition of the stationary phase. Here we recommend conditions that will improve the overall PCA pattern, in order to better characterise their occurrence in future environmental monitoring and exposure assessment.