The Duration and Intensity of High Frequency Alternating Current Influences the Degree and Recovery of Nerve Conduction Block.

IEEE transactions on bio-medical engineering(2024)

引用 0|浏览2
暂无评分
摘要
OBJECTIVE:The purpose of this paper is to investigate the persistence of nerve blockade beyond the duration of applying high frequency alternating current (HFAC) to thinly myelinated and non-myelinated fibers, also termed a "carry-over effect". METHODS:In this study, we used electrically-evoked compound action potentials from isolated rat vagus nerves to assess the influence of 5 kHz HFAC amplitude and duration on the degree of the carry-over effect. Current amplitudes from 1-10 mA and 5 kHz durations from 10-120 seconds were tested. RESULTS:By testing 20 different combinations of 5 kHz amplitude and duration, we found a significant interaction between 5 kHz amplitude and duration on influencing the carry-over effect. CONCLUSION:The degree of carry-over effect was dependent on 5 kHz amplitude, as well as duration. SIGNIFICANCE:Utilizing the carry-over effect may be useful in designing energy efficient nerve blocking algorithms for the treatment of diseases influenced by nerve activity.