Abstract
Culture pH is a critical process parameter during CHO cell bioreactor operations that is key for proper cell growth, protein production, and maintaining the critical quality attributes of a monoclonal antibody drug substance. The traditional means of measuring pH in bioreactors is with an electrochemical probe that can withstand and maintain accuracy through repeated sterilization cycles. An alternative technique for measuring pH is an optical sensor composed of a fluorescent dye that is sensitive to the hydrogen ion concentration. In this work we explore singleuse electrochemical and single-use optical pH sensors in stirred-tank and rocking bioreactors, respectively, to understand how their overall performance compares to traditional electrochemical probes in benchtop glass stirred tank bioreactors. We found that the single-use optical pH sensors were generally less accurate than the electrochemical probes, especially in detecting large pH drifts from the setpoint. The single-use electrochemical probes were increasingly accurate as pH was increased from <7.0 to 7.5 but tended to decrease in accuracy as the batch age increased. In conclusion, single-use pH sensors offer a convenient means to measure pH during an upstream bioprocess, but the limitations of these sensors should be built into process control such that deviations in process pH, and consequently potential fluctuations in product quality, can be avoided.
More
Translated text
Key words
Chinese hamster ovary cells,CHO,Process control,pH control,pH sensor,Monoclonal antibodies,Bioreactors,Bioprocessing,Biomanufacturing,Single -use technology
AI Read Science
Must-Reading Tree
Example
Generate MRT to find the research sequence of this paper
Chat Paper
Summary is being generated by the instructions you defined