Neutron dosimetry with a pair of TLDs for the Elekta Precise medical linac and the evaluation of optimum moderator thickness for the conversion of fast to thermal neutrons

Marziyeh Behmadi, Sara Mohammadi, Mohammad Ehsan Ravari, Aghil Mohammadi,Mahdy Ebrahimi Loushab,Mohammad Taghi Bahreyni Toossi,Mitra Ghergherehchi

NUCLEAR ENGINEERING AND TECHNOLOGY(2024)

引用 0|浏览0
暂无评分
摘要
Introduction: In this study, TLD 600 and TLD 700 pairs were used to measure the neutron dose of Elekta Precise medical linac. To this end, the optimum moderate thickness for the conversion of fast to thermal neutrons were evaluated. Materials and methods: Am-241-Be and Cf-252 sources were simulated to calculate the optimum thicknesses of the moderator for the conversion of maximum fast neutrons (FN) into thermal neutrons (TN). Pair TLDs were used to measure F&TN doses for three different field sizes at four depths of the medical linac. Results: The maximum thickness of the moderator was optimized at 6 cm. The measurement results demonstrated that the TN dose increased with the expansion of field size and depth. The FN dose, which was converted TN, exhibits behaviors comparable to the TN due to its nature. Conclusion: This study presents the optimum thickness for the moderator to convert FN into TN and measure F&TN using TLDs.
更多
查看译文
关键词
Radiotherapy,Neutron contamination,Optimum moderate thickness,Am-241-Be and(252)Cf sources,Monte Carlo simulation,TLD