ART and AutoART ECAP measurements and cochlear nerve anatomy as predictors in adult cochlear implant recipients

European archives of oto-rhino-laryngology : official journal of the European Federation of Oto-Rhino-Laryngological Societies (EUFOS) : affiliated with the German Society for Oto-Rhino-Laryngology - Head and Neck Surgery(2024)

引用 0|浏览3
暂无评分
摘要
Purpose The purpose of this retrospective study is to compare the results of electrically evoked compound action potential (ECAP) measurements using automatic auditory response telemetry (AutoART) with those obtained by ART in adults. The study also aimed to evaluate the predictive value of intraoperative ART and AutoART ECAPs for speech intelligibility (SI) and hearing success (HS), and to determine if cochlear nerve (CN) cross-sectional area (CSA) obtained preoperatively by magnetic resonance imaging (MRI) scans could predict ART and AutoART ECAPs and SI and HS outcome. Methods The study analyzed and correlated ART and AutoART ECAP thresholds at electrodes E2, E6, and E10, as well as averaged ECAP thresholds over electrodes E1–E12, using data from 32 implants. Correlations were also examined for ART and AutoART ECAP slopes. In addition, averaged ART and AutoART ECAP thresholds and slopes over all 12 electrodes for each participant were correlated with CN CSA measured from MRI sequences. SI of the monosyllabic Freiburg Speech Test at 65 dB sound pressure level was examined along with averaged ART and AutoART thresholds and slopes over all 12 electrodes. A parallel analysis was performed for HS, derived from the difference between baseline and 6-month SI. Finally, correlations between CN CSA and SI, as well as CN CSA and HS were examined. Results The results of the study showed a significant positive correlation between ART and AutoART ECAP thresholds and as well as slopes for E2, E6, E10 and averaged thresholds and slopes of E1–E12. However, no significant correlation was observed between ART and AutoART averaged ECAP thresholds and slopes and either SI and HS or CN CSA. Furthermore, no significant correlation was found between CN CSA and SI and HS. Conclusion While AutoART is a reliable and safe program for measuring ECAPs in adults, the study found no preoperative prognostic information on intraoperative ECAP results using parameters extracted from current MRI sequences or pre-/intraoperative information on subsequent hearing outcome using ECAP and CN CSA.
更多
查看译文
关键词
Cochlear implant,Cochlear nerve,Electrically evoked compound action potentials,Magnetic resonance imaging,Prognostic factors
AI 理解论文
溯源树
样例
生成溯源树,研究论文发展脉络
Chat Paper
正在生成论文摘要