Miniaturized, high numerical aperture confocal fluorescence detection enhanced with pyroelectric droplet accumulation for sub-attomole analyte diagnosis

BIOMEDICAL OPTICS EXPRESS(2023)

引用 0|浏览2
暂无评分
摘要
To meet the growing demand for early fatal disease screening among large populations, current fluorescence detection instruments aiming at point-of-care diagnosis have the tendency to be low cost and high sensitivity, with a high potential for the analysis of low-volume, multiplex analytes with easy operation. In this work, we present the development of a miniaturized, high numerical aperture confocal fluorescence scanner for sub-micro-liter fluid diagnosis. It is enhanced with high-rate analyte accumulation using a pyroelectro-hydrodynamic dispensing system for generating tiny, stable sample droplets. The simplified confocal fluorescence scanner (numerical aperture 0.79, working distance 7.3 mm) uses merely off-the-shelf mass-production optical components. Experimental results show that it can achieve a high-sensitive, cost-efficient detection for sub-micro-liter, low-abundant (0.04 mu L, 0.67 attomoles) fluid diagnosis, promising for point-of-care diagnosis.Published by Optica Publishing Group under the terms of the Creative Commons Attribution 4.0 License. Further distribution of this work must maintain attribution to the author(s) and the published article's title, journal citation, and DOI.
更多
查看译文
关键词
confocal fluorescence detection,pyroelectric droplet accumulation,high numerical aperture,sub-attomole