Superior bone fixation of conical compared with hemispherical trapezial cup design: an experimental radiostereometry study
Journal of Experimental Orthopaedics(2023)
摘要
The most used cup designs for trapeziometacarpal (TMC) arthroplasty are of hemispherical and conical geometrical shape. Using a validated pig bone model, we compared the bone fixation using radiostereometry (RSA). Twenty saddle-shaped pig forefoot bones were prepared with insertion of bone markers and reaming. Hemispherical Type T cups (Beznoska, Kladno, Czech Republic) (N = 10) and conical Moovis cups (Stryker, Pusignan, France) (N = 10) of 9-mm diameter were inserted press-fit. The bones were fixed in cement blocks for stability, and the cups were loaded in a motorized test stand. First, a low-pressure cyclic load test (0—150N) with 130 compression cycles was performed. Next, a push-in test of progressive loads with 50N increments (range: 150–900N) was applied until a visual change in cup position appeared. Cup migration was evaluated with RSA after every new load application. Cup failure was defined as total translation > 0.5 mm between two load applications. Both cup types tolerated a compression load of 450 N without failure. Beyond this load level, the total translation cup migration of mean 0.20 mm (95
更多查看译文
关键词
Trapeziometacarpal arthroplasty,Trapeziometacarpal osteoarthritis,Cup fixation,Radiostereometry,Conical cup,Hemispherical cup