A look-up-table development to facilitate CT simulation of MR-Linac treatment

PHYSICS & IMAGING IN RADIATION ONCOLOGY(2024)

引用 0|浏览10
暂无评分
摘要
While current MR-Linac (MRL) treatment workflows utilize a large table overlay during CT simulation to convert indexing between the two machines, we developed a look-up-table (LUT) as an alternative approach. After populating the LUT, index conversion factors were verified at three separate table locations. The resultant root-mean-square isocenter shifts on the MRL were 0.04/0.08 cm, 0.08/0.07 cm, and 0.09/0.08 cm with/without using the table overlay during simulation in the lateral, longitudinal, and vertical directions, respectively, which is within registration tolerance. Clinical implementation of the LUT has resulted in a more efficient MRL treatment workflow while maintaining accurate patient setup.