Leukocyte differential based on an imaging and impedance flow cytometry of microfluidics coupled with deep neural networks.

Cytometry. Part A : the journal of the International Society for Analytical Cytology(2024)

引用 0|浏览14
暂无评分
摘要
The differential of leukocytes functions as the first indicator in clinical examinations. However, microscopic examinations suffered from key limitations of low throughputs in classifying leukocytes while commercially available hematology analyzers failed to provide quantitative accuracies in leukocyte differentials. A home-developed imaging and impedance flow cytometry of microfluidics was used to capture fluorescent images and impedance variations of single cells traveling through constrictional microchannels. Convolutional and recurrent neural networks were adopted for data processing and feature extractions, which were then fused by a support vector machine to realize the four-part differential of leukocytes. The classification accuracies of the four-part leukocyte differential were quantified as 95.4% based on fluorescent images plus the convolutional neural network, 90.3% based on impedance variations plus the recurrent neural network, and 99.3% on the basis of fluorescent images, impedance variations, and deep neural networks. Based on single-cell fluorescent imaging and impedance variations coupled with deep neural networks, the four-part leukocyte differential can be realized with almost 100% accuracy.