Variation Minimization Based Electrocardiogram Artifacts Removal for Local Field Potentials From Neurostimulator

IEEE TRANSACTIONS ON NEURAL SYSTEMS AND REHABILITATION ENGINEERING(2024)

引用 0|浏览0
暂无评分
摘要
Local field potential (LFP) recorded by sensing-enabled neurostimulators provided chronic observation of deep brain activities for the research of brain disorders. However, the contamination from the electrocardiogram (ECG) deteriorated the extraction of effective information from LFP. This study proposed a novel algorithm based on minimizing the variance combining template subtraction to improve the performance of ECG artifact removal for LFP. Four patients with implanted electrodes were recruited, and eight real LFP records were collected from their left and right hemispheres, respectively. The results showed that the proposed method improved the accuracy of artifact peak detection in LFP, and the subsequent signal quality after template subtraction compared to the traditional Pan-Tompkins (PT) method. The outcome of this study benefited the LFP-based brain research, promoting the application of sensing-enabled neurostimulators in more areas.
更多
查看译文
关键词
Local field potential (LFP),template subtraction,artifact removal,sensing-enabled neurostimulator
AI 理解论文
溯源树
样例
生成溯源树,研究论文发展脉络
Chat Paper
正在生成论文摘要