In‐vivo quality assurance of dynamic tumor tracking (DTT) for liver SABR using EPID images

Journal of Applied Clinical Medical Physics(2023)

引用 1|浏览2
暂无评分
摘要
Abstract Purpose To assess dynamic tumor tracking (DTT) target localization uncertainty for in‐vivo marker‐based stereotactic ablative radiotherapy (SABR) treatments of the liver using electronic‐portal‐imaging‐device (EPID) images. The Planning Target Volume (PTV) margin contribution for DTT is estimated. Methods Phantom and patient EPID images were acquired during non‐coplanar 3DCRT‐DTT delivered on a Vero4DRT linac. A chain‐code algorithm was applied to detect Multileaf Collimator (MLC)‐defined radiation field edges. Gold‐seed markers were detected using a connected neighbor algorithm. For each EPID image, the absolute differences between the measured center‐of‐mass (COM) of the markers relative to the aperture‐center (Tracking Error, (E T )) was reported in pan, tilt, and 2D‐vector directions at the isocenter‐plane. Phantom study An acrylic cube phantom implanted with gold‐seed markers was irradiated with non‐coplanar 3DCRT‐DTT beams and EPID images collected. Patient Study : Eight liver SABR patients were treated with non‐coplanar 3DCRT‐DTT beams. All patients had three to four implanted gold‐markers. In‐vivo EPID images were analyzed. Results Phantom Study : On the 125 EPID images collected, 100% of the markers were identified. The average ± SD of E T were 0.24 ± 0.21, 0.47 ± 0.38, and 0.58 ± 0.37 mm in pan, tilt and 2D directions, respectively. Patient Study : Of the 1430 EPID patient images acquired, 78% had detectable markers. Over all patients, the average ± SD of E T was 0.33 ± 0.41 mm in pan, 0.63 ± 0.75 mm in tilt and 0.77 ± 0.80 mm in 2D directions The random 2D‐error, σ, for all patients was 0.79 mm and the systematic 2D‐error, Σ, was 0.20 mm. Using the Van Herk margin formula 1.1 mm planning target margin can represent the marker based DTT uncertainty. Conclusions Marker‐based DTT uncertainty can be evaluated in‐vivo on a field‐by‐field basis using EPID images. This information can contribute to PTV margin calculations for DTT.
更多
查看译文
关键词
dynamic tumor tracking,liver sabr,epid images,quality assurance
AI 理解论文
溯源树
样例
生成溯源树,研究论文发展脉络
Chat Paper
正在生成论文摘要