Pulsed-field ablation in patients with permanent pacemaker or implantable cardioverter defibrillator: Is it safe?

Europace(2023)

引用 0|浏览6
暂无评分
摘要
Abstract Funding Acknowledgements Type of funding sources: None. Aims Pulse field ablation (PFA) is a novel catheter ablation technology with safety benefits due to its tissue selectivity. The risk of physical damage or electromagnetic interference (EMI) of PFA with cardiac implantable electronic devices (CIEDs) is unknown. The aim of our study was to assess the risk of PFA for CIEDs. Methods PFA lesions (45 per CIED) were applied to CIEDs (<5cm from the lead tip and <15cm from the generator). All devices were checked before and after PFA application for sensing and pacing parameters as well as integrity of shock circuits in ICDs using a heart simulator. Devices were interrogated for any spontaneous reprogramming, mode switching or other EMI effects. Results In total, 44 CIEDs were tested (16 pacemaker, 21 ICDs, 7 CRT-P/D) with 1,980 PFA applications. There was no change in device settings, functionality, electrical parameters and there was no visible damage to the CIEDs or leads. The risk on a device-based analysis is 0/44 (95% CI 0%-8%) and on a PFA pulse-based analysis is 0/1,980 (95% CI 0%-0.2%). Clinically relevant EMI appeared with oversensing and pacing inhibition, but not tachycardia detection. Conclusions Bipolar PFA appears safe and does not result in damage to CIEDs or leads. Clinically relevant EMI does occur, but appropriate peri-procedural programming may mitigate this. In-vivo studies are needed to confirm our findings.
更多
查看译文
关键词
implantable cardioverter defibrillator,implantable cardioverter,permanent pacemaker,pulsed-field
AI 理解论文
溯源树
样例
生成溯源树,研究论文发展脉络
Chat Paper
正在生成论文摘要