Orthopedic implants affect the electric field induced by switching gradients in MRI

Magnetic Resonance in Medicine(2023)

引用 0|浏览9
暂无评分
摘要
Abstract Purpose To investigate whether the risk of peripheral nerve stimulation increases in the presence of bulky metallic prostheses implanted in a patient's body. Methods A computational tool was used to calculate the electric field (E‐field) induced in a realistic human model due to the action of gradient fields. The calculations were performed both on the original version of the anatomical model and on a version modified through “virtual surgery” to incorporate knee, hip, and shoulder prostheses. Five exam positions within a body gradient coil and one position using a head gradient coil were simulated, subjecting the human model to the readout gradient from an EPI sequence. The induced E‐field in models with and without prostheses was compared, focusing on the nerves and all other tissues (both including and excluding the bones from the analysis). Results In the nerves, the most pronounced increase in the E‐field (+24%) was observed around the knee implant during an abdominal MRI ( Y axis readout). When extending the analysis to encompass all tissues (excluding bones), the greatest amplification (+360%) occurred around the knee implant during pelvic MRI ( Z axis readout). Notable increases in E‐field peaks were also identified around the shoulder and hip implants in multiple scenarios. Conclusion Based on the presented results, further investigations aimed at quantifying the threshold of nerve stimulation in the presence of bulky implants are desirable.
更多
查看译文
关键词
<scp>mri</scp>,electric field