Improvement in the precision of capillary refill time measurements for diagnosing hypovolemic status

Journal of Biomechanical Science and Engineering(2023)

引用 0|浏览0
暂无评分
摘要
The capillary refill time (CRT) is the time that elapses before the fingernail color returns to red from white after releasing a load applied to the nail. We aimed to estimate hypovolemic status using the CRT. However, the precision of CRT is low. The first step toward improving precision was estimation of the coefficient of variance (CV) via repeated measurements. The best and next-best indices of the CRT change rates in the hypovolemic state were 0.19 and 0.48 respectively, and the corresponding target CVs were ≤ 0.10 and 0.24. The CRTs of subjects were measured by imaging nails during loading and unloading and fitting of an exponential function to the curve of intensity changes. The CRT was 0.77 ± 0.35 s (n = 108), which is comparable to previous studies. However, the CV was 0.37 ± 0.13, which is higher than even the second-best index. Thus, we used only the green color channels of nail images to generate algorithms to define the optimal start time of the CRT calculations; this provided a lower CV (0.31 ± 0.12). Finally, we investigated the effects of fingertip temperature and load release timing on CRT during a cardiac cycle. Correction of the temperature effect with a care on a phase during the cardiac cycle to trigger data collection reduced CV to 0.15 ± 0.11, which cleared the second-best index. Thus, our protocol improves CRT precision; it may be possible to estimate hypovolemic status using the CRT.
更多
查看译文
关键词
capillary refill time measurements,time measurements
AI 理解论文
溯源树
样例
生成溯源树,研究论文发展脉络
Chat Paper
正在生成论文摘要