Hydrus microstent for the treatment of primary open angle glaucoma: overview of its safety and efficacy

Expert Review of Medical Devices(2023)

引用 0|浏览0
暂无评分
摘要
Minimally invasive glaucoma surgeries (MIGS) are now a consolidated reality in many surgical units. The Hydrus Microstent is one of several MIGS devices bypassing trabecular outflow and had excellent results over the years. This article aims to review the key features of the Hydrus Microstent in terms of design, efficacy, and safety.The present review analyses the main characteristics of the device by evaluating the technical and physical details of its functioning. The evidence that supports a clinical decision summarizes the most influential clinical trials and the most accurate systematic reviews.The Hydrus device has been extensively studied regarding biocompatibility and outflow potential. The subsequent clinical studies have been well-built and proved that the device effectively reduces intraocular pressure (IOP) and the eyedrop load. The device covers almost a quarter of Schlemm's canal circumference, offering at least two advantages: cannulating the Schlemm's canal provides evidence that the device has been implanted correctly; covering a larger area potentially allows to target multiple collector channels or at least areas of active outflow. This scaffold may prove more effective in naïve patients or subjects who used antiglaucoma eyedrops for a limited period, as the prolonged use of hypotonic medications has been associated with the surgical failure of ab interno microhook trabeculotomy.