Optimizing cardiac rehabilitation by pacemaker reprogramming. Can we immediately improve the exercise capacity?

Balneo and PRM Research Journal(2023)

引用 0|浏览4
暂无评分
摘要
Background: Previous studies have demonstrated a direct correlation between pacemaker programming mode and exercise capacity on a long-term follow-up. In the present study we aimed to test whether reprogramming of the pacemaker can produce immediate effects on patients' exercise capacity. Methods: The exercise capacity of 33 patients wearing dual-chamber pacemakers was evaluated by cardiopulmonary exercise test at 24-hour intervals in two programming modes: initially in the single-chamber ventricular pacing mode with a fixed 60bpm rate, and in the second phase in dual chamber pacing mode with exercise adaptable rate, up to 130bpm. Results: The maximum O2 consumption (VO2 max: 12.82±2.70 mL/kg/min vs 14.52±3.25 mL/kg/min, p=0.02), the O2 consumption at the time of the anaerobic threshold (10.76±2.25 mL/kg/min vs 12.30±2.84 mL/kg/min, p=0.008), the duration of exercise (456.76±116.85 seconds vs 510.57±129.56 seconds, p=0.04) and the maximum workload (84.39±17.89W vs 96.36±24.09W, p=0.01) were significantly lower when the pacemakers were programmed in the single chamber ventricular fixed-rate pacing mode compared to the dual chamber rate adaptive pacing mode. Conclusion: Cardiac pacemaker reprogramming is able to exert immediate effects on exercise capacity. Dual chamber pacing with adaptable rate during physical exercise is superior to single chamber fixed rate pacing
更多
查看译文
关键词
pacemaker reprogramming,cardiac rehabilitation,exercise capacity
AI 理解论文