Development of Colloidal Gold Test Strips for the Detection of Oxamyl Residues in Tobacco.

Studies in health technology and informatics(2023)

Cited 0|Views11
No score
Abstract
In this study, monoclonal antibodies against oxamyl were prepared, and colloidal gold immunochromatography was used to design a rapid test strip product for the detection of oxamyl in tobacco with high specificity, accuracy and stability without cross-reactivity to commonly used tobacco fungicides based on the optimization of conditions such as pH value of diluent, diluent dosage, concentration of antibody marker, type of confining solution and complex solution. 5 The results of five samples of post-harvest ready-to-bake tobacco and first-harvest tobacco were consistent with the gas chromatographic method, which proved the reliability of the test strips. The limits of detection for the post-harvest and first-harvest tobacco samples were 0.1 mg/kg, and the test strips developed in this study are suitable for mass testing in tobacco laboratories with good application prospects because of their short detection time, simple pre-treatment and detection methods.