A low-cost, portable 32-channel EIT system with four rings based on AFE4300 for body composition analysis

Joel Escobar Fernandez, Cristian Martinez Lopez,Victor Mosquera Leyton

HARDWAREX(2023)

引用 0|浏览0
暂无评分
摘要
A proposed low-cost, portable, 32-channel (4 rings of 8-channel) Electrical Impedance Tomography (EIT) system based on the AFE4300 analog front-end for body composition measurement. Each ring allows obtaining the conductivity distribution of 4 cross sections, 4 cm apart; to analyze the behavior of conductivity in a volume. The switching of the 4 rings and the current injection and voltage measurement patterns are done with three Texas Instruments 74HC4067 multiplexers, which are managed by an ESP32 board. The proposed system has an average signal-to-noise ratio of 74.71 dB and a frame rate of 50 fps. The sensitivity tests to impedance and volume changes consisted of introducing 4 tubes of different diameters (2 steel and 2 polyvinyl chloride) into a tank with saline solution; then conductivity distribution images were generated in 4 cross-sections of the tank, using the algorithms Gauss-Newton and Noser. Finally, the global impedance index (GI) is calculated to estimate the volume of each tube inside the tank. The results show that the proposed system is highly sensitive to impedance and volume changes, being a promising system for monitoring tissues, and fluids biological.
更多
查看译文
关键词
Electrical impedance tomography,Bioimpedance,Image reconstruction,Global impedance