谷歌Chrome浏览器插件
订阅小程序
在清言上使用

Daytime hypoglycemic episodes during the use of an advanced hybrid closed loop system

Diabetes research and clinical practice(2023)

引用 0|浏览13
暂无评分
摘要
Aims: The use of advanced hybrid closed loop systems is spreading due to the beneficial effects on glycometabolic control obtained in patients with type 1 diabetes. However, hypoglycemic episodes can be sometimes a matter of concern. We aim to compare the hypoglycemic risk of an advanced hybrid closed loop system and a predictive low glucose suspend sensor augmented pump. Methods: In this retrospective three months observational study, we included 30 patients using Medtronic MinimedTM 780G advanced hybrid closed loop system and 30 patients using a Medtronic MinimedTM predictive low glucose suspend sensor augmented pump. Results: The advanced hybrid closed loop system reduced the time spent above 180 mg/dL threshold and increased the time in range as compared to the predictive low glucose suspend. No severe hypoglycemia occurred in both groups and no differences were observed in the percentage of time spent below 70 mg/dl and 54 mg/dl glucose threshold. Nevertheless, more hypoglycemic episodes were recorded during daytime, but not in night-time, with the use of the advanced hybrid closed loop system. Conclusions: Our results confirmed the general improvement of glycemic outcomes obtained with the advanced hybrid closed loop system; however more hypoglycemic episodes during daytime were evident.
更多
查看译文
关键词
Type 1 diabetes,Advanced hybrid closed loop system,Hypoglycemia,Nighttime hypoglycemia,Daytime hypoglycemia
AI 理解论文
溯源树
样例
生成溯源树,研究论文发展脉络
Chat Paper
正在生成论文摘要