A case study for in-house method validation of gas chromatography technique using class-1 calibration gas mixtures for greenhouse gases monitoring

ACCREDITATION AND QUALITY ASSURANCE(2023)

引用 0|浏览1
暂无评分
摘要
In this study, analytical method validation has been done for the measurement of carbon dioxide/nitrogen (CO 2 /N 2 ) and methane/nitrogen (CH 4 /N 2 ) calibration gas mixtures using gas chromatography with flame ionization detector (GC-FID). Class-I calibration gas mixtures (CGMs) of CO 2 (500 µmol mol −1 to 1100 µmol mol −1 ) and CH 4 (2 µmol mol −1 to 130 µmol mol −1 ) used in method validation process has been prepared gravimetrically following ISO 6142-1. All prepared gas mixtures have expanded uncertainty 1 % at coverage factor ( k ) of 2 with 95 % confidence. The following parameters are chosen for this case study which include selectivity, accuracy, precision, linearity, limit of detection (LOD), limit of quantification (LOQ), robustness, stability, and uncertainty. Different statistical approaches are taken into consideration for each parameter assessment. The results indicate that GC-FID is selective for CO 2 and CH 4 . CGMs represent good repeatability and reproducibility having percentage relative deviation < 1 % among measurements. A good linear behaviour was observed for CGMs of CH 4 and CO 2 on basis of least square regression with R 2 value 0.9995 and 1, respectively. LOD and LOQ for CH 4 are calculated 0.47 and 1.59 µmol mol −1 based on signal-to-noise ratio by taking its lowest concentration of 2.9 µmol mol −1 . The in-house validated method for GC-FID using CGMs for the measurement of greenhouse gases (CO 2 and CH 4 ) is found to be precise, accurate and fit for purpose.
更多
查看译文
关键词
Greenhouse gases,GC-FID,Method validation,Calibration gas mixture
AI 理解论文
溯源树
样例
生成溯源树,研究论文发展脉络
Chat Paper
正在生成论文摘要