Purpose. This in vitro study compared the vertical marginal discrepancy (MD) and absolute marginal discrepancy (AMD) between the zirconia substructure and titanium (Ti)-base abutment before and after cementation. Material and methods. An intraoral scanner (CEREC Omnicam scanner; Dentsply Sirona) was used to scan the dental model of missing tooth 36 and obtain a three-dimensional virtual model. It was designed and milled for 25 zirconia superstructures (inCoris ZI meso F2 L, Dentsply Sirona). The samples were cemented to Ti bases. The MD and AMD were measured for each sample by using a scanning electron microscope (SEM) before and after cementation. Results. The overall mean values of MD measurements before and after cementation were 14.12 mu m and 20.28 mu m, respectively. The overall mean values of AMD measurements prior to and after substructure cementation on the Ti base were 54.85 mu m and 116.0 mu m, respectively. ANOVA revealed significant differences in MD (p = 0.001) and AMD (p < 0.001) before and after cementation. Conclusions. Significant differences were observed in MD and AMD values before and after cementation. The MD and AMD values were significantly lower before cementation than after cementation. The mean MD and AMD values were clinically acceptable.